The SECULIFE Ox one FingerSIMâ„¢ test system
makes it possible for technical medical personnel
to evaluate pulse oximeters and sensor functions
under three simulated light absorption conditions.
Typical values of 80%, 90% and 97% SpO2 can be
simulated.

With the help of this new simulation method, interaction
between the sensor’s sensitivity to light and